  • Measuring range: 0.05 µS/cm to 1 mS/cm
  • Process pressure up to 16 bar
  • Cell constant K = 0.01 or 0.1
Basic type: 202924

Conductive two-electrode conductivity sensor

These sensors are used in conjunction with conductivity transmitters to define the electrolytic conductivity in liquids. The materials in the sensors are physiologically harmless and meet FDA requirements.

Diverse process connections; robust design type; pharmaceutical version incl. ASTM certificate

General technical data
Measuring range0.05 μS/cm to 1 mS/cm
Electrode materialStainless steel (1.4571 or 1.4435); titanium
Process connectionThread G 1/2 A, thread G 3/4 A, thread G 1 A, thread 1/2“-14 NPT, thread 3/4“-14 NPT, taper socket for union nut DN 25, DIN 11851 (dairy pipe fitting), clamp DN 25
Process pressure16 bar at 25 °C or 9 bar at 60 °C; high-temperature version: 40 bar at 25 °C or 17 bar at 200 °C
Operating temperatureUp to 135 °C (200 °C as optional extra)
Cell constantK = 0.01 or 0.1

Purified water and highly-purified water; boiler feed water; chip manufacturing; ion exchangers and reverse osmosis plants; high-temperature applications