dTRANS T01

  • Pt100 internal
  • Sampling rate: > 1 measurement per second
  • Output signal 4 to 20 mA
Category: Brand:

Download Documents

Basic type: 707010

Two-wire transmitter

The two-wire transmitter acquires the temperature through an RTD temperature probe or a thermocouple. For RTD temperature probes, the probe can be connected in a two-wire circuit, three-wire circuit, or four-wire circuit. The dTRANS T01 Ex is designed for mounting in a potentially explosive area.

General technical data
Ambient temperature -40 to +85 °C
Voltage supply DC 11.5 to 30 V
RTD temperature probe input Pt100, Pt500, Pt1000 (Junior only Pt100/Pt1000), two-wire/three-wire/four-wire circuit
Installation In terminal head, type B
Installation On mounting rail/DIN rail 35 × 7.5 mm
Special features Customer-specific linearization
Output 4 to 20 mA
Configuration Via PC interface, HART® version via HART® modem
Output 4 to 20 mA (invertible)
NAMUR compliant NE21
Approval ATEX/IECEx
Calibration accuracy ± 0.05 %
Galvanic isolation 3.75 kV, 2 kV, 1 kV
Configuration Via HART® modem
Voltage supply DC 11.5 to 30 V (two-wire transmitter)
Galvanic isolation 2 kV
RTD temperature probe input Pt100, Pt500, Pt1000, two-wire/three-wire/four-wire circuit
Thermocouple input J, K (more upon request)
Thermocouple input L, J, U, T, K, E, N, S, R, B, D, C (Junior only J, K, N, S, R)
Ambient temperature -25 to +70 °C